If your IMA system has a Relay Logic Control sequence panel, the contacts in the mechanical relays could be degrading. There are several solutions: change the position of the relays on the sequence panel and see if the system will complete the cycle, or replace the old relays with new ones. Another possibility is that either the Head Closed switch failed to close during the cycle, causing it to be interrupted. Check the function of the switches.